Table 7-5: SDI-12 commands and values returned
1
SDI-12 command
2
Values returned Units
Response
time
aM! or aMC!
1. Address (0 through 9)
2. Relative wind direction
3. Relative wind speed
4. Corrected wind direction
5. Status (4-digit code); see
Sensor status code (p. 23)
1. n/a
2. °
3. m/s
4. °
5. n/a
3 s
aM1! or aMC1!
1. Address (0 through 9)
2. Temperature
3. Relative humidity
4. Dew point
5. Barometric pressure
6. Status (4-digit code); see
Sensor status code (p. 23)
7. Wind chill
8. Heat index
9. Air density
10. Wet bulb temperature
1. n/a
2. °C
3. %
4. °C
5. hPa
6. n/a
7. °C
8. °C
9. kg/m
2
10. °C
2 s
aM2! or
aMC2!
1. Address (0 through 9)
2. Relative wind direction
3. Relative wind speed
4. Status (4-digit code); see
Sensor status code (p. 23)
1. n/a
2. °
3. m/s
4. n/a
3 s
aM3! or
aMC3!
1. Address (0 through 9)
2. Precipitation intensity
3. Total precipitation
4. Status (4-digit code); see
Sensor status code (p. 23)
1. n/a
2. mm/hr
3. mm
4. n/a
3 s
?! Returns the SDI-12 Address
1
Not all outputs are available for all sensors.
2
a is the SDI-12 address. In the SDI12Recorder() CRBasic instruction, the command parameter does not
include the SDI-12 address because the address is a separate parameter.
MetSENS-Series Compact Weather Sensors 19