9
④ If you set sample size to 10, place 10 samples on the platter.
⑤ Press the SET key. The display shows unit weight and then shows the number of samples. If the
display shows “LACK(◀ )”, the counting result may not correct. (LCD version : Lack of Unit lamp is On) If
the display shows “ Err9”, it is impossible to count samples because the unit weight is too small. (Refer to Table 1)
Model
ED 3kg ED 6kg ED 15kg ED 30kg
Range of Lack
1 g≥Lack 2 g≥Lack 5 g≥Lack 10 g≥Lack
Range of Low
0.5g≥ Err9 1g≥ Err9 2 g≥ Err9 5g≥ Err9
< Table 1 >
⑥ Place parts on the platter, then the display shows the number
of parts. You can also use tare function in counting mode
4. Percent Weighing Mode
OR
① Press the MODE key until the display shows “PErSEt”. If you
have set reference sample before, the display shows “0.0”.
② If you want to use container,
place it and press the TARE key.