CEFALY Dual - Problem #05 - the Intensity Feels too Low

CEFALY Dual
101 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
71

Possible Cause:
Once you’re accustomed to the sensaons experienced while using the CEFALY
DUAL you may feel desensized to its lower-strength pulses at the beginning of a
session.
Soluon:
10 seconds aer you’ve selected your session, you may press and hold the device
buon to quickly ramp up the intensity to your desired strength. When you release
the device buon, the intensity will stabilize and remain constant. Should you press
and hold the device buon again, the intensity will increase again. The maximum
intensity is reached aer pressing and holding the device buon for 30 consecuve
seconds. You should increase the intensity to a tolerable level, but you should stop
before the sensaon becomes intolerable or painful.