EasyManua.ls Logo

cefar IFC - Device Functionality and Controls; Starting a Treatment Session

cefar IFC
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6
Turns the stimulator on and off.
Selects programs 1–7.
Programming button for P7.
Used to confirm settings
for P7.
Increases amplitude.
Press and hold the
button to increase the
amplitude continuously.
Note: Be very careful
when increasing the
amplitude.
Reduce amplitude. Press
and hold the button to
decrease the amplitude
continuously. Also deacti-
vates the keylock.
FUNCTION
Indicates that the keylock
is activated.
Shows remaining run
time for the program.
Shows the selected
amplitude in mA.
1. Connect the electrodes to the cables.
Wash and dry the skin well before
applying the electrodes. If the skin is
hairy, it may need to be shaved.
2. Place the electrodes in the area to
be treated. See the Programs and
Electrode Placement chapter.
3. Switch the stimulator on with .
4. Select a program using P.
5. Increase the amplitude with
.
6. When the program time runs out,
the stimulation switches off and the
timer symbol blinks. Switch the sti-
mulator off with
. If you do not
switch the stimulator off, it turns off
automatically after 5 minutes to spare
the batteries.
STARTING TREATMENT
Number of program
phases.
2470-200509_CEFAR_IFC_manual_inlaga_outline.indd 2005-09-14, 15:186

Table of Contents