41
05 Relevant G-code commands
G92 Xnnn Ynnn Znnn
G92 sets the current posion of the printhead to the specied X, Y and Z coordinates. If no values are
given, the posion is assumed to be 0, 0, 0 (this will also change the Z posion).
Tx
Tx switches to printhead (x), where 0, 1 and 2 designate printheads 1, 2 and 3.
M771 Tx Pyy
M771 turns on the heater in printhead Tx at temperature Pyy, where yy is a temperature between 30
and 65 degrees Celsius.
M750 Tx Py Dz
Extrude from printhead x at pressure y for z milliseconds.
M751 Tx
Stop extrusion from printhead x.
M2065 Tx Sy
Set valve open me of printhead x to y microseconds.
M2067 Tx Sy
Set valve cycle me of printhead x to y microseconds.
M400
M400 command causes all G-code processing to pause and wait unl the printer nishes all movement
(i.e., the planner queue is empty).