EasyManua.ls Logo

Center 306 - Thermocouple Probe Specifications; TP-K01 Probe Details

Center 306
11 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Data Logger Instruction Manual
6
bit3: 1Indicates the LCD is displaying time.
bit4: no use
bit5: 1HOLD, 0not HOLD
bit6: 1LOW BATTERY , 0BATTERY NORMAL
bit7: 1°C 0°F
3
th
BYTE:
bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0
Auto Power Off memory full resolution sign OL resolution sign OL
bit0: 1→T1 is OL, 0not OL
bit1: 1→T1 value is minus, 0→T1 value is plus.
bit2: 14
th
byte and 5
th
byte represent #### , 04
th
byte and 5
th
byte represent ###.#
bit3: 1→T2 is OL, 0not OL
bit4: 1→T2 value is minus, 0→T2 value is plus.
bit5: 1→8
th
byte and 9
th
byte represent #### , 0→8
th
byte and 9
th
byte represent ###.#
bit6: 1Memory is full. 0Memory is not full.
bit7: 1Auto power off enabled. 0Auto power off disabled.
4
th
BYTE: first two BCD code of T1 value.
5
th
BYTE:last two BCD code of T1 value
6
th
BYTE:
If bit3 of 2
nd
BYTE =0 : first two BCD code of T1-T2 value.
If bit3 of 2
nd
BYTE =1 : two BCD code of month.
7
th
BYTE:
If bit3 of 2
nd
BYTE =0 : last two BCD code of T1-T2 value.
If bit3 of 2
nd
BYTE =1 : two BCD code of day.
8
th
BYTE:
If bit3 of 2
nd
BYTE =0 : first two BCD code of T2 value.
If bit3 of 2
nd
BYTE =1 : two BCD code of hour.
9
th
BYTE:
If bit3 of 2
nd
BYTE =0 : last two BCD code of T2 value.
If bit3 of 2
nd
BYTE =1 : two BCD code of minute.
10
th
BYTE: end byte, it value is 3, 1
nd
and 10
th
are used to check frame error.
Appendix: Thermo couple probe specification
Model Range Tolerances Description
TP-K01
Bead probe
-50 to 200
-58 to 392
±2.2 or ±0.75
(±3.6 or ±0.75)
with Teflon tape insulation Maximum
insulating temperature : 260
TP-K01:
probe for general condition measurements, especially for
complex and hard to reach places.

Related product manuals