Data Logger Instruction Manual
6
bit3: 1→Indicates the LCD is displaying time.
bit4: no use
bit5: 1→HOLD, 0→not HOLD
bit6: 1→LOW BATTERY , 0→BATTERY NORMAL
bit7: 1→°C 0→°F
3
th
BYTE:
bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0
Auto Power Off memory full resolution sign OL resolution sign OL
bit0: 1→T1 is OL, 0→not OL
bit1: 1→T1 value is minus, 0→T1 value is plus.
bit2: 1→4
th
byte and 5
th
byte represent #### , 0→4
th
byte and 5
th
byte represent ###.#
bit3: 1→T2 is OL, 0→not OL
bit4: 1→T2 value is minus, 0→T2 value is plus.
bit5: 1→8
th
byte and 9
th
byte represent #### , 0→8
th
byte and 9
th
byte represent ###.#
bit6: 1→Memory is full. 0→Memory is not full.
bit7: 1→Auto power off enabled. 0→Auto power off disabled.
4
th
BYTE: first two BCD code of T1 value.
5
th
BYTE:last two BCD code of T1 value
6
th
BYTE:
If bit3 of 2
nd
BYTE =0 : first two BCD code of T1-T2 value.
If bit3 of 2
nd
BYTE =1 : two BCD code of month.
7
th
BYTE:
If bit3 of 2
nd
BYTE =0 : last two BCD code of T1-T2 value.
If bit3 of 2
nd
BYTE =1 : two BCD code of day.
8
th
BYTE:
If bit3 of 2
nd
BYTE =0 : first two BCD code of T2 value.
If bit3 of 2
nd
BYTE =1 : two BCD code of hour.
9
th
BYTE:
If bit3 of 2
nd
BYTE =0 : last two BCD code of T2 value.
If bit3 of 2
nd
BYTE =1 : two BCD code of minute.
10
th
BYTE: end byte, it value is 3, 1
nd
and 10
th
are used to check frame error.
Appendix: Thermo couple probe specification
Model Range Tolerances Description
TP-K01
Bead probe
-50℃ to 200℃
-58℉ to 392℉
±2.2℃ or ±0.75%
(±3.6℉ or ±0.75%)
with Teflon tape insulation Maximum
insulating temperature : 260℃
TP-K01:
probe for general condition measurements, especially for