Manual
Laboratory autoclave
Autoclave using steam for the treatment of material and laboratory processes e.g.
sterilization
CertoClav MultiControl 2
ENGLISH
German version page 25

Table of Contents

Other manuals for CertoClav MultiControl 2

Questions and Answers

  • T
    Tara GreerAug 14, 2025
    What does 'heater error' mean on CertoClav Laboratory Equipment?
    • Z
      Zachary PatelAug 14, 2025
      A 'heater error' message on your CertoClav laboratory equipment indicates that the maximum allowed temperature has been exceeded (143°C) or the target temperature was not reached. Please verify that the temperature sensor is functioning correctly and that the autoclave chamber has sufficient water.
  • K
    Kimberly SantosAug 17, 2025
    What to do if CertoClav Laboratory Equipment displays 'power loss'?
    • M
      Mr. Steven BowmanAug 17, 2025
      If your CertoClav laboratory equipment displays 'power loss', ensure the fuse holder is tightly assembled and the power cable is correctly mounted.
  • C
    Christopher BoydAug 20, 2025
    How to fix a CertoClav Laboratory Equipment with display background light turned off permanently?
    • S
      shanesimmonsAug 21, 2025
      If the display background light on your CertoClav laboratory equipment is permanently off, check if the power switch is on. Replace both 5mm 10A slow-acting fuses located at the black power connector's shelf.
  • J
    Joe WestAug 25, 2025
    What causes a 'sensor error' on CertoClav Laboratory Equipment?
    • E
      Emily BeasleyAug 25, 2025
      A 'sensor error' on your CertoClav laboratory equipment usually indicates an issue with the temperature sensor. This often means the sensor is not plugged into the controller, or there may be a cable break.
  • G
    Grace KramerAug 27, 2025
    What does 'delta-T error' mean on CertoClav Laboratory Equipment?
    • M
      martin69Aug 28, 2025
      A 'delta-T error' on your CertoClav laboratory equipment means the allowed temperature range of T-1/+2°C has been exceeded during the holding time. Check that the temperature cable is correctly plugged and screwed into the controller and that the autoclave chamber is not out of water.
  • B
    Billy MartinAug 31, 2025
    Why is water dropping out of the gasket of the lid of my CertoClav Laboratory Equipment?
    • A
      Amanda ChavezAug 31, 2025
      Water may drop from the lid gasket of your CertoClav laboratory equipment during heat-up as the gasket compresses. If it persists after heat-up, remove the gasket and clean it with distilled water. If the problem continues, contact CertoSupport to order a new gasket.
  • S
    suzannesmallSep 3, 2025
    What does 'Program stopped' mean on CertoClav Laboratory Equipment?
    • R
      Ryan LaneSep 3, 2025
      The 'Program stopped' message on your CertoClav laboratory equipment indicates that the user has manually stopped the program.
  • L
    leroywhiteSep 6, 2025
    How to resolve 'profiler error' on CertoClav Laboratory Equipment?
    • B
      Barbara SmithSep 7, 2025
      To resolve a 'profiler error' on your CertoClav laboratory equipment, navigate to Settings -> Factory reset and press the OK button. This should address software issues.
  • L
    lschneiderSep 10, 2025
    How to adjust the display contrast on CertoClav Laboratory Equipment?
    • K
      Kara HernandezSep 10, 2025
      To adjust the display contrast on your CertoClav laboratory equipment, navigate to Settings -> Contrast and press the OK button until the contrast is satisfactory.

CertoClav MultiControl 2 Specifications

General IconGeneral
BrandCertoClav
ModelMultiControl 2
CategoryLaboratory Equipment
LanguageEnglish