3D ActiveTrac
®
32
OPERATION
TOUCHSCREEN CONTROL PANEL TCP CONTINUED
Important Points: If the table reaches its full excursion (fully
opens) before maximum force has been reached, the force the
patient actually receives will be less than what is displayed on
the TCP. To prevent this from occurring, ensure that:
•Thetableisintheclosedandlockedpositionduringset-up.
Unlock before pressing Start.
•TheBeltsand/orHarnessesfitsnuglyagainstbareskinand
do not slip.
•Monitorexcursionidentificationmarkingstoassurefull
excursion has not been reached.
•Ifpatientrepositioningisrequired,youshouldStoprather
than Pause the unit, so that you can close and lock the table
during repositioning. You cannot close the unit during Pause
mode unless you Pause during a Rest Time.
•Ifyourepositionedthepatient,remembertotakeupthe
slack in the belts by pulling on the Pelvic Harness straps at
the lumbar end of the table.