Legend 2 - Service Manual DOMESTIC 7
GENERAL WARNINGS AND PRECAUTIONS
This device should be used only under the continued supervision of a physician or licensed practitioner.
Contaminated sponges, electrodes, lead wires, and gel can lead to infection.
Use of electrode on multiple patients can lead to infection.
Do not apply electro stimulation treatment during bath, shower, sauna,..
DO NOT operate the Intelect® Legend 2 within the vicinity or environment of an ultrasonic diathermy system.
DO NOT operate the Intelect® Legend 2 within the vicinity or environment of any microware and RF shortwave
diathermy system.
DO NOT operate this unit in an environment where other devices are being used that intentionally radiate
electromagnetic energy in an unshielded manner.
Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate properly when
electrical stimulation is in use.
Simultaneous connection of a PATIENT to a high frequency surgical ME EQUIPMENT may result in burns at the
site of the STIMULATOR electrodes and possible damage to the STIMULATOR.
Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the
Intelect® Legend 2, including cables specified by the manufacturer. Otherwise, degradation of the performance
of this equipment could result.
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Battery replacement by inadequately trained personnel could result in fire or explosion. Please read carefully the
battery replacement instructions in the Legend 2 IFU before attempting to replace the battery.
Device is designed to comply with electromagnetic safety standards. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with instructions for use, may cause
harmful interference to other devices in the vicinity. Harmful interference to other devices can be determined by
turning this equipment on and off. Try to correct the interference using one or more of the following:
» Reorient or relocate the receiving device
» Increase the separation between the equipment
» Connect the equipment to an outlet on a different circuit from that to which the other device(s) are connected »
Consult your authorized ENOVIS dealer for help.
Disconnect the system from the power source before attempting any maintenance, installation, removal, or
replacement procedures to prevent electrical shock and possible damage to system.
The Intelect® Legend 2 may be susceptible to Electro-Static Discharge (ESD) at greater than ±6 kV when first
grasping the Ultrasound applicator. In the event of such a discharge, the Intelect® Legend 2 may display a
permanent error. The Intelect® Legend 2 will terminate all active outputs (stim, ultrasound,), automatically
place the unit in a safe state.