EasyManua.ls Logo

Chroma 66201 - The Commands of the Power Meter; Standard Commands

Chroma 66201
107 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Digital Power Meter 66200 Series User’s Manual
5.8
5.8.1
The Commands of the Power Meter
Standard Commands
*CLS 66201 | 66202
Description: This command clears the status byte register and the event registers.
Setting Syntax: *CLS<PMT>
Setting Parameters: none
Query Syntax: none
Response Parameters: none
Example: none
ESE 66201 | 66202
Description: This command sets the standard event status enable register.
This command programs the Standard Event register bits. If one or more of the
enabled events of the Standard Event register is set, the ESB of Status Byte
Register is set too.
Bit Configuration of Standard Event Status Enabled Register
Bit Position 7 6 5 4 3 2 1 0
Bit Name PON - - - CME EXE - - - QYE - - - - - -
CME = Command error QYE = Query error
EXE = Execution error PON = Power-on
Setting Syntax: *ESE<space><NR1><PMT>
Setting Parameters: <NR1>, 0 ~ 255
Query Syntax: *ESE?<PMT>
Response Parameters: <NR1>, 0 ~ 255
Header on: *ESE<space><NR1><RMT>
Header off: <NR1><RMT>
Example: none
*ESR? 66201 | 66202
Description: This command reads out the contents of the standard event status register
(SESR).
Setting Syntax: none
Setting Parameters: none
Query Syntax: *ESR?<PMT>
Response Parameters: <NR1>, 0 ~ 255
Header on: <NR1><RMT>
Header off: <NR1><RMT>
Example: none
5-8

Other manuals for Chroma 66201

Related product manuals