R
Ryan MurphyAug 14, 2025
How to fix output current over high (general) on Chroma Test Equipment?
- CChase EllisAug 14, 2025
If your Chroma Test Equipment shows an output current that is generally too high, try widening the I LIMIT setting.
How to fix output current over high (general) on Chroma Test Equipment?
If your Chroma Test Equipment shows an output current that is generally too high, try widening the I LIMIT setting.
How to fix DC Bus voltage over high on Chroma 69225-100-4?
If you are experiencing issues with high DC Bus voltage on your Chroma Test Equipment, make sure to calibrate it periodically and observe if the 691101 action is normal.
What to do if DC Bus voltage is over low on Chroma 69225-100-4 Test Equipment?
To resolve a low DC Bus voltage issue with your Chroma Test Equipment, calibration periodically is need. Additionally, observe if the 691101 action is normal.
What to do if local voltage is over high on Chroma 69225-100-4?
If your Chroma Test Equipment shows a high local voltage, check if the output terminal wiring is loosened and verify that the BMS is operating normally.
What to do if output current transient is extremely high (fast) on Chroma Test Equipment?
If your Chroma Test Equipment displays an extremely high transient output current, check if the output terminal wiring is loose and ensure the BMS is operating normally.
What to do if local connected reversely on Chroma 69225-100-4?
If your Chroma Test Equipment indicates the local connection is reversed, check to ensure the output terminal connects to the positive/negative terminal exactly.
What to do if RMT connected reversely on Chroma 69225-100-4 Test Equipment?
If your Chroma Test Equipment reports that the remote (RMT) connection is reversed, check if the sensor wire connects to the positive/negative terminal correctly.
How to fix wire loss too large (or RMT wire off) on Chroma 69225-100-4 Test Equipment?
If your Chroma Test Equipment shows excessive wire loss, check the output wire path for any abnormal or high impedance objects. Also, verify that the output and RMT wires are not loose and fix them if necessary.
How to prevent 692 inner over-temperature on Chroma 69225-100-4 Test Equipment?
The Chroma Test Equipment should be operated under ambient temperature 0°C~ 40°C.
Brand | Chroma |
---|---|
Model | 69225-100-4 |
Category | Test Equipment |
Language | English |
Details the operational modes and protection functions of the tester.
Outlines essential preparation steps before operating the instrument.
Details the voltage, frequency, current, and power requirements.
Guides on connecting output terminals for charging/discharging.
Introduces the troubleshooting procedures for operational issues.
Lists common failure causes and their corresponding resolutions.