ELECTOROMAGNETIC COMPATIBILITY INFORMATION
WARNING
• Portable RF communications equipment should be used no closer than 30 cm (12
inches) to any part of the [CHUD514/517], including cables specied.
• Use of this equipment adjacent to or stacked with other equipment should be avoided.
• Use of accessories and options other than those specied (other than CITIZEN
original parts) could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment.
Guidance and manufacturer’s declaration - electromagnetic emissions
The [CHUD514/517] is intended for use in the electromagnetic environment specied below. The
customer or the user of the [CHUD514/517] should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The [CHUD514/517] uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The [CHUD514/517] is suitable for use
in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/ icker
emissions IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration - electromagnetic immunity
The [CHUD514/517] is intended for use in the electromagnetic environment specied below.
The customer or the user of the [CHUD514/517] should assure that it is used in such an
environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for power
supply lines
±1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV
line to line
±2 kV
line to earth
±1 kV
line to line
±2 kV
line to earth
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
IEC 61000-4-11
0% U
T
0.5 cycle
0% U
T
0.5 cycle
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
[CHUD514/517] requires continued
operation during power mains
interruptions, it is recommended that
the [CHUD514/517] be powered from
an uninterruptible power supply or a
battery.
0% U
T
1 cycle
0% U
T
1 cycle
70% U
T
25/30 cycle
70% U
T
25/30 cycle
0% U
T
250/300 cycle
0% U
T
250/300 cycle
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic elds should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Note: U
T
is the A.C. mains voltage prior to application of the test level.
CHUD514_517_1802_ENES.indb 15 2018/12/05 9:34