EasyManua.ls Logo

Citizen CTB-300 - Scale Calibration; Calibration Procedure

Citizen CTB-300
31 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CTB serial bench scale service manual
13
SECTION 10 CALIBRATION
Turn the power off. During the counting from 9 to 0 press the FUNC/C key.
The display will show "CAL " for a few seconds. While it is showing "CAL " press
the UINT , PRINT and TARE keys in sequence to enter the Calibration section.
The display will show "unLoAd".
Remove any weight from the platform. Press the ZERO/ENTER key.
The display will show "LoAd". Place the calibration weight on the scale. Press the
ZERO/ENTER key.
If the calibration weight is right, the display will show “PASS”, If the calibration weight
is unapt, the display will show “FAIL H” or “FAIL L. the calibration is not
acceptable.
If the calibration is acceptable the display will return to normal. If an error message is
shown try calibration again as a disturbance may have prevented a successful
calibration.
If the problem persists then contact your dealer.
SECTION 11 TECHNICAL PARAMETERS
Press F key when normal weighing mode, display shows F0 H-L”, press TARE key
until display shows P ROG”, press ZERO key, display shows PIN”, You can press
unit PRINT TARE key to enter setting mode, press Tare key to select parameter,
press Zero key to sure, press Print key to escape. Before you set parameters, please
short the K1.
FUNCTION
SUB-FUNCTION DESCRIPTION

Related product manuals