EasyManua.ls Logo

Clearfield YOURx - Page 32

Clearfield YOURx
37 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Direct: 763.476.6866 • National: 800.422.2537 • www.SeeCleareld.com • techsupport@clfd.net
32
YOURx Multi-Purpose Terminal
Installation Manual
_________________________________________________________
Manual 021023 REV A - Jan 2020
Cleaning the Ferrule
Lightly moisten the ber optic swab (2.5mm/38542F or 1.25mm/38040)
by spotting a small amount (about 1”) of Electro-Wash PX or
Electro-Wash MX pen onto the QbE. Hold the swab, 1 side down to the
wetted area and hold for a count of 1-2-3-4-5. (Figure 4)
Figure 4
Figure 5
Figure 6
Cleaning the Mate Through an Adapter AND the Adapter Itself
Lightly moisten the ber optic swab (2.5mm/38542F or 1.25mm/38040)
by spotting a small amount (about 1”) of Electro-Wash PX or
Electro-Wash MX pen onto the QbE. Hold the tip of the swab onto the
wetted area and hold for a count of 1-2-3-4-5.
Insert the swab into the adapter to the connector, press lightly against
the connector, twist 2-3 times, remove and discard.
Dry with a second dry swab.
Inspect, repeat cleaning if necessary, and test for signal strength.
Use additional swabs to clean inside the actual adapter. Moisten
swab, like above, and insert through hole and remove while twisting.
(Figure 6)
Insert swab into side of ferrule, wet side to the ceramic ferrule and
circle around 2-3 times and remove. Turn swab to dry side and repeat.
(Figure 5)