5 Installation and Set Up
T34L™ Operator Manual
NOTE: Replace the battery once every two years in certified service centers.
NOTE: Charge the unit with dedicated T34L
TM
charger only.
Battery Test
When setting up the pump on a patient always check that there is enough battery charge remaining to
cover the infusion duration. To do so follow one of the procedures below:
Check the Start Up Screen, displayed on power up, indicating % battery life remaining
Press INFO key when the pump is stopped and select Battery Level, press YES to view battery
meter
During infusion press the INFO key twice to check
Battery Level, in all cases verify sufficient battery charge
is available to complete the current program. If not,
charge the battery.
Sufficient battery charge may vary depending on your
setting, infusion rate, number of times the information
options/screen backlight are used.
Battery Level
85%
Empty Full