Syringe pump with motor driven linear actuator, pulsed motion
(60 pulses per mm)
2 ml to 60 ml (most commonly used manufacturers, see Default Syringe section
for more details)
100-1500mmHg configurable (10mmHG increments).
Max. Actuator force 50N (5 Kgf).
Rechargeable Li-Ion 1800 mAh
c. 15 hours at 125ml/hr rate
Graphic Display 128 x 64 pixels with back light (8 keys) & dual color
operation LED
When a problem is detected, the T34L
TM
displays the following alarm messages,
sounds an audible alarm and the red LED lights:
Syringe not Recognized
Occlusion
End Program
Near End/Nearly Complete
End Syringe Travel
Low Battery
End Battery
Pump Paused
Lock Mode
Syringe Displaced
Check Syringe Placement
Complies with: EN 60601-1 (Medical Electrical Equipment Safety),
IEC 60601-2-24 (Infusion pumps and controllers), IEC 62304 (Medical device
software – Software life cycle processes)
Manufacture in accordance to ISO 9001:2008 and ISO 13485:2003. CE marked
(In accordance with the Medical Devices Directive 93/42/EEC)
Environmental
Specifications
Non Operating Conditions (Transportation and Storage): -
Temperature: -25oC to 50 oC (-13 oF to + 122 oF)
Humidity: 5 % to 100% R.H., non-condensing
Air pressure: 48kPa to 110kPa
Operating Conditions: The system may not meet all performance specifications if
operated outside of the following conditions: -
Temperature: +15oC to +45 oC (+59 oF to + 113 oF)
Humidity: 5 % to 90% R.H. at +40 oC, non-condensing
Air pressure: 70kPa to 110kPa