User manuel
-22
* Adjustable timer: 10~60 minutes
* Adjustable intensity: 20 level
* Multiple program selection: 24 programs
* Rechargeable lithium battery: DC 3.7V
* Wide voltage design: AC 100-240V 50/60Hz
4. Unit description
A. Turn on/off the Device
▲Note: Do not turn on the unit until all electrodes and lead wires are properly attached.
1) Press the On/Off button for 3s, the LCD will light up and the device will turn on. Press the On/Off button for 3s, the display
will go blank and the d
evice will turn off.
2) The device will turn off automatically after the cure time has elapsed.
3) In an emergency you may also pull the connector(s) of electrode cable from the device to remove the electrode pads.
▲Note: To prevent unpleasant electric shocks, never remove the pads while it is still turned on.
B. Display
The touchable LCD screen displays treatment remaining time, battery power, program icons, output port, current intensity,
selected intensity and lock sta
te. The default value of TENS and Muscle Stimulator is Program 1, Output A, Intensity 00, Timer
20min.
*Output channel
The TENS and Muscle Stimulator has 2 output channel: Channel A and Channel B. Press “A/B” button to select Output
Channel.
*Programs
The TENS and Muscle Stimulator offers in totally 24 different preset treatment programs. All parameters are pre-programmed,
except for the intensity and cure time.
Press program icons on the touchable LCD to select the progr
am. Once you select a program, the corresponding program
number flashes per 0.5 second. The LCD display screen shows below. Note: Please refer to chapter 3 “User Instruction” to
learn how to choose the right treatment program and time.
*Intensity
Once you select a program, gradually increase the intensity by pressing the “+” button or clicking “+” icons, or decrease the
intensity by pressing the “–” button or clicking “-” icons, until you get the desired intensity level.
▲Note: If you change a program during this treatment, the intensity level will revert to the default value of “00” to protect
your safety. Channel A and channel B are independent, and you can set their intensity separately.
▲Note: You will
feel the intensity increase or decrease as you select the intensity level. You can use this as a guide to select a
level that is comfortable for you. At the precondition of acceptance, the intensity should be chosen as high as possible for the
manual