Chapter 16
TROUBLESHOOTING
ENGLISH
Excessive movement
Over range
Saturated signal
The signal extent is too big owing to
the arm or body moving or other rea-
sons when measuring
Keep arm, body still, measure again
Time out It takes too much time
Abnormal Phenomenons Causes Solutions
BP measurement values too high or
too low
Cuff is not connected correctly. Correctly connect cuff
(refer to Chapter 10)
Talk or move arms when measuring Keep quiet and restart a measurement
The turnup clothing presses the arm Take off the clothing which presses the
arm, and restart a measurement
No pressure Cuff leakage Buy a new cuff
The cuff windpipe is not correctly con-
nected with cuff
Correctly connect
Cuff is not inated Stop using the device and contact us
Cuff deates in short time Loose cuff Correctly apply cuff
It can not carry on measurement when press the measurement button Switch on the power once again and
restart a measurement