88
Name SpO2 Probe (Accessory Separate Sale)
Measurement Range SpO2 Measuring Range: 0%~100%;
Pulse Rate Measuring Range: 30 bpm~250 bpm;
Resolution
SpO2 1%
PR 1bpm
Measurement Accuracy
SpO2 70%~100% ±2%
0%~69% undened
PR ±2 bpm or ±2% (select larger)
Measurement Performance In Weak Filling Condition
Pulse-lling ratio : 0.4% SpO2 error ±4%
Pulse rate error ±2 bpm or ±2% (select larger)
Optical Sensor
Red light wavelength is 660 nm, 6.65 mW
Infrared wavelength is 880 nm, 6.75 mW
Chapter 20
SpO2 SPECIFICATION
(This chapter is only suitable for European Union market)