Overpressure protection adult mode 295±5 mmHg (39,33±0,67 kPa)
pediatric mode 240±5 mmHg (32±0,67 kPa)
neonatal mode 145±5 mmHg (19,33±0,67 kPa)
Ination adult 160±5 mmHg (21,33±0,67 kPa)
pediatric 120±5 mmHg (16±0,67 kPa)
neonatal 70±5 mmHg (9,33±0,67 kPa)
Prompt Range adult mode SYS: 40~270 mmHg - DIA: 10~215 mmHg
pediatric mode SYS: 40~200 mmHg - DIA: 10~150 mmHg
neonatal mode SYS: 40~135 mmHg - DIA: 10~100 mmHg
Resolution Pressure: 1mmHg (0.133 kPa)
Accuracy Static pressure: ±3 mmHg (±0.4 kPa)
Error The BP Value of the device is equivalence with the measurement value of Stethosco-
py. The error meets the requests in the ANSI/AAMI SP-10:2002+A1:2003 +A2:2006
Operating temperature/ humidity +5ºC~40ºC 15% RH~85% RH (Non-condensing)
Transport Transport by general vehicle or according to the order contract, avoid pounded,
shake and splash by rain and snow in transportation.
Storage Temperature: -20ºC~+55ºC; Relative humidity: ≤95%; No corrosive gas and drafty.
Atmospheric pressure 700 hPa~1060 hPa
Power supply 4 "AA" alkaline batteries, AC Adapter(AC, 100 V-240 V, optional)
Rated current ≤ 600 mA
Battery life When the temperature is 23ºC, limb circumference is 270 mm, the measured blood
pressure is normal, 4 "AA" alkaline batteries cab be used about 300 times.
Dimensions 130(L) × 110(P) × 80(A) mm
Unit Weight 300 gram (without batteries)