Guidance and manufacturers declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specied below. The customer or the user of the
device should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3 Vrms
150 kHz
to 80 MHz
3 V/m
80 MHz
to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment should be used
no closer to any part of the device, including cables, than the recom-
mended separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the rec-
ommended separation distance in metres (m).
Field strengths from xed RF transmitters, as determined by an elec-
tromagnetic site survey,
a
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
3.5
V
1
P
d=
3.5
E
1
P
d=
7
E
1
P
d=
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Guidance and manufacturer’s declaration electromagnetic immunity for EQUIPMENT and SYSTEMS that are
not LIFE-SUPPORTING