Principles of the Procedure: The CONTOUR NEXT ONE blood
glucose test is based on measurement of electrical current
caused by the reaction of the glucose with the reagents on the
electrode of the test strip. The blood sample is drawn into the tip
of the test strip through capillary action. Glucose in the sample
reacts with FAD glucose dehydrogenase (FAD-GDH) and the
mediator. Electrons are generated, producing a current that is
proportional to the glucose in the sample. After the reaction
time, the glucose concentration in the sample is displayed. No
calculation by the user is required.
Comparison Options: The C
ONTOUR NEXT ONE system is
designed for use with capillary whole blood. Comparison with a
laboratory method must be done simultaneously with aliquots of
the same sample.
NOTE: Glucose concentrations drop rapidly due to glycolysis
(approximately 5%–7% per hour).