Manuale di Istruzione
Instruction Manual
21/05/2012 58-E4800 Rev. 4 EN
The instrument is powered by two rechargeable batteries type “R6 AA size” 1,2V –
2400 mAh min.
With fully charged and efficient batteries the instrument can be used for approximately 14
hours with a pulse rate of 1 Hz.
Working with a higher pulse rate and/or using the back lighting will reduce the operation
time.
NOTE:
Remember that batteries have a small current of auto discharge, therefore it is
recommended to recharge them periodically (for example once a month) even if the
instrument is not being used.
The led on the instrument panel will show the charge status of the batteries as follows:
led off = batteries charged
led flashing = limited autonomy
with battery charger connected, led flashing = charging completed.
To maintain the batteries efficient, we recommend that they are only recharged when
fully discharged avoiding intermediate or incomplete charging.
When recharging is complete the display will show:
Fig. 3-3
The led on the front panel will flash.
It takes approximately 14 hours to completely recharge of the batteries.
During recharging it is recommended that the battery charger is not
disconnected/reconnected.
The batteries can be recharged up to approximately 1000 times if recharged correctly.