Manuale di Istruzioni
Instruction Manual
58-E4800 Rev. 4 EN 21/05/2012
21
Inversely, by pressing the arrow up key it is possible to adjust the gain: from High
(H) to Low (L).
Normally use the High (H) gain.
Fig. 3-8
The Low (L) gain should only be used when examining small samples (e.g. 10 cm) and
when an oscilloscope is connected (see chapter 3.4); Indeed, this selection should be
made when in normal conditions (with gain high) the wave shape is square rather than a
sine wave. In this case, reducing the gain to Low (L) a sine wave will be re-obtained.
NOTE:
Setting Low (L) in all other cases can result in incorrect readings.