Output Receptacle Replacement
8-22 ForceTriad Energy Platform Service Manual
Output Receptacle Replacement
Equipment required:
Phillips screwdriver
1. Remove the front-panel assembly; see Removing the Front Panel on page 8-2.
2. Remove the Steering-Relay PCBA; see Steering-Relay PCBA Replacement on page 8-
16.
Monopolar 1 Receptacle
a. Remove the three screws securing the Monopolar 1 receptacle to the front panel.
b. Remove the scanner shroud.
c. Install the new Monopolar 1 receptacle and scanner shrouds in the front panel
using the three screws.
Monopolar 2 Receptacle
a. Remove the three screws securing the Monopolar 2 receptacle to the front panel.
b. Remove the scanner shroud.
c. Install the new Monopolar 2 receptacle and scanner shrouds in the front panel
using the three screws.
Warning
Electric Shock Hazard To allow stored energy to dissipate after power is disconnected, wait at
least five minutes before replacing parts.
Caution
The system contains electrostatic-sensitive components. When repairing the system, work at a
static-control workstation. Wear a grounding strap when handling electrostatic-sensitive
components, except when working on an energized system. Handle PCBAs by their non-
conductive edges. Use an antistatic container for transport of electrostatic-sensitive components
and PCBAs.
Notice
Perform all the steps including the recalibration listed below. Failure to recalibrate the system after
replacing components may result in the system becoming inoperable.