Sterilization and cleaning
-Steam autoclave is recommended.
-Sterilization requirement after using as noted below.
-Autoclave Procedure:
-Scrub dirt and debris from the handpiece,and wipe clean with alcohic
cotton.Do not use a wire brush(Fig.18.4)
-Lubricate with SPRAY or Lubricant oil.
-Lubrication.
-Insert into an autoclave pouch.Seal the pouch.
-Autoclavable up to a max.135°C(EX.)Autoclave for 20 min. at 121°C or
15 min.at 132°C(Fig.18.2/18.3).
Heat sterilized 135°C Heat sterilized 135°C
the tubing contra angle for implant
2 8
Heat sterilized 135°C
Fig.18.3
Fig.18.2