User Manual for Wrist Oximeter
-25-
F. Preset limits:
SpO
2
Low limit setting: 85%
Pulse Rate limit setting range:
Upper limit: 120bpm
Lower limit: 50bpm
G. Update rate:
6 seconds moving average for SpO
2
and 8 beats
average for Pulse Rate readings.
H. Perfusion Index Display
Range: 0.2%~20%
I. Performance under low perfusion condition
The accuracy of SpO
2
and PR measurement still
meets the specification described above when the
modulation amplitude is as low as 0.6%.
J. Resistance to interference of surrounding light:
The difference between the SpO
2
value measured in
the condition of indoor natural light and that of
darkroom is less than ±1%.
K. Resistance to 50Hz /60Hz interference:
SpO
2
and PR are precise which have been tested by
BIO-TEK pulse oximeter simulator.