sirius+ | IFU327EN00 - 01/2021
INSTRUCTIONS FOR USE
This document is the property of
CSO Costruzione Strumenti Oftalmici srl.
Any reproduction, even partial, it is prohibited.
39/56
4.4 HOW TO TURN ON THE DEVICE
Read the instructions for use before using the application software.
1
Press the activation switch of the power supply unit on ON.
2
Turn on the PC.
3
Launch the Phoenix application software.
4
Wait until the main screen of the application software is
shown.
5
If this is the first time you’re starting the device, or after a long
period of inactivity, calibration shall be performed. Follow the
instructions given in paragraph “
How to perform device
calibration” on page 39.
4.4.1 HOW TO PERFORM DEVICE CALIBRATION
Calibration shall be performed when starting the device for the first
time or after a long period of inactivity of the latter.
The procedure shall be carried out in a dark room so as to simulate
the environmental conditions of a standard capture procedure.
Follow the instructions about calibrating the device sirius+ which ar
e
given in the Phoenix application software handbook.
Particular attention shall be paid while performing such procedure. It
is important to check device stability before startin
g with the
procedure.
The calibration is essential to obtain precise measurements.
1
Make sure the calibration tool is clean and undamaged. If
needed, clean it using a soft cloth.
Do not use solvents or diluents to clean the calibration tool.
2
Place the calibration tool on the chin rest.