INSTALLER’S INSTRUCTIONS
RELEASE 7.03
Group 2 lockup, LED # 6 of group 2 blinks and machine emits 3 beeps. To cancel the alarm turn the main
switch of the machine off and back on. This alarm is triggered when the electronic board senses that the
readings of the temperature probe in group 2 are out of range.
Temperature probe Group 3
Group 3 lockup, LED # 6 of group 3 blinks and machine emits 3 beeps. To cancel the alarm turn the main
switch of the machine off and back on. This alarm is triggered when the electronic board senses that the
readings of the temperature probe in group 3 are out of range.
Temperature probe Group 4
Group 4 lockup, LED # 6 of group 4 blinks and machine emits 3 beeps. To cancel the alarm turn the main
switch of the machine off and back on. This alarm is triggered when the electronic board senses that
the readings of the temperature probe in group 4 are out of range
Flowmeter Group 1
LED of dispensing button being used blinks. To stop dispensing, press button P3 of group 1. To cancel
the alarm turn the main switch of the machine off and back on. This alarm is triggered when the electronic
board cannot receive any readings from flowmeter of group 1 during coffee dispensing.
Flowmeter Group 2
LED of dispensing button being used blinks. To stop dispensing, press button P3 of group 2. To cancel
the alarm turn the main switch of the machine off and back on. This alarm is triggered when the electronic
board cannot receive any readings from flowmeter of group 2 during coffee dispensing.
Flowmeter Group 3
LED of dispensing button being used blinks. To stop dispensing, press button P3 of group 3. To cancel
the alarm turn the main switch of the machine off and back on. This alarm is triggered when the electronic
board cannot receive any readings from flowmeter of group 3 during coffee dispensing.
Flowmeter Group 4
LED of dispensing button being used blinks. To stop dispensing, press button P3 of group 4. To cancel
the alarm turn the main switch of the machine off and back on. This alarm is triggered when the electronic
board cannot receive any readings from flowmeter of group 4 during coffee dispensing.
Timeout on Group 1
LED # 3 on group 1 blinks and machine emits 1 beep. To cancel alarm press button P3 of group 1. This
alarm is triggered when button P3 of group 1 is used to dispense coffee continuously for over 4 minutes.
Timeout on Group 2
LED # 3 on group 2 blinks and machine emits 1 beep. To cancel alarm press button P3 of group 1. This
alarm is triggered when button P3 of group 1 is used to dispense coffee continuously for over 4 minutes.
Timeout on Group 3
LED # 3 on group 3 blinks and machine emits 1 beep. To cancel alarm press button P3 of group 3. This
alarm is triggered when button P3 of group 3 is used to dispense coffee continuously for over 4 minutes.
Timeout on Group 4
LED # 3 on group 4 blinks and machine emits 1 beep. To cancel alarm press button P3 of group 4. This
alarm is triggered when button P3 of group 4 is used to dispense coffee continuously for over 4 minutes.
PARAMETERS RANGE
Set Boiler 110 – 125°C step 0.1°C
Set Group 1 80 – 110°C step 0.1°C
Set Group 2 80 - 110°C step 0.1°C
Set Group 3 80 - 110°C step 0.1°C
Set Group 4 80 - 110°C step 0.1°C
Boiler Delta T 1 - 5°C step C
Extra-set 0 - 5°C step C
On Coffee 0 - 20 sec
Time-out autofill 1 - 6 min