ERROR
MESSAGES
CAUSE
SOLUTION
RX1 TUBEHEAD
IS NOT CONNECTED OR IS OUT OF ORDER
CALL THE “ASSISTANCE SERVICE”
RX2 TUBEHEAD
IS NOT CONNECTED OR IS OUT OF ORDER
CALL THE “ASSISTANCE SERVICE”
CORRUPTED EEPROM DATA CALL THE “ASSISTANCE SERVICE”
EEPROM DATA NOT SAVED PROPERLY CALL THE “ASSISTANCE SERVICE”
LINE VOLTAGE VALUE NOT INCLUDED
WITHIN THE SET LIMITS
CALL THE “ASSISTANCE SERVICE”
LINE VOLTAGE VALUE NOT INCLUDED
WITHIN THE ±15% NOMINAL VALUE
CALL THE “ASSISTANCE SERVICE”
THE X-RAY KEY
ALWAYS SEEMS TO BE PRESSED
MAKE SURE IT IS NOT JAMMED
ANOMALY IN THE CONTROL PANEL CALL THE “ASSISTANCE SERVICE”
THE EXPOSURE
HAS BEEN PREMATURELY INTERRUPTED
KEEP THE X-RAY KEY
PRESSED TILL THE END OF THE EXPOSURE
ANOMALY IN THE TRIAC/RELAY CALL THE “ASSISTANCE SERVICE”
ANOMALY IN THE ELECTRONIC CIRCUIT CALL THE “ASSISTANCE SERVICE”
ANOMALY IN THE CONTROL CIRCUIT CALL THE “ASSISTANCE SERVICE”
INCORRECT DIP-SWITCH CONFIGURATION CALL THE “ASSISTANCE SERVICE”
THE CONTROL BUTTON DOES NOT CORRESPOND
TO THE SELECTED TUBEHEAD
CALL THE “ASSISTANCE SERVICE”
THE TUBEHEAD
DOES NOT WORK PROPERLY
CALL THE “ASSISTANCE SERVICE”
THE TUBEHEAD
IS NOT IN THE CORRECT MODE
CALL THE “ASSISTANCE SERVICE”
THE TUBEHEAD
HAS NOT COMPLETED THE EXPOSURE
REPEAT THE EXPOSURE
CALL THE “ASSISTANCE SERVICE”
PROBLEM IN THE FREQUENCY
OR REGULATION
CALL THE “ASSISTANCE SERVICE”
THE TUBEHEAD
IS NOT CALIBRATED
CARRY OUT WITH THE CALIBRATION
CALL THE “ASSISTANCE SERVICE”
EEPROM DATA NOT SAVED PROPERLY CALL THE “ASSISTANCE SERVICE”
CORRUPTED EEPROM DATA CALL THE “ASSISTANCE SERVICE”
OVERVOLTAGE ERROR CALL THE “ASSISTANCE SERVICE”
ANODE VOLTAGE OUT OF TOLERANCE CALL THE “ASSISTANCE SERVICE”
ANODE CURRENT OUT OF TOLERANCE CALL THE “ASSISTANCE SERVICE”
CONTROL CONNECTOR CALL THE “ASSISTANCE SERVICE”
PROBLEM IN THE REFERENCE VOLTAGE CALL THE “ASSISTANCE SERVICE”
MAJOR ERROR
ALL FUNCTIONS ARE DISABLED
CALL THE “ASSISTANCE SERVICE”