15.1 Automatic Cell Cleaning Unit Fault Finding
Pneumatic
Cleaner not
operating
Air connections missing
or wrong.
Sort out air connections. Air
connections are Push-In type. Press
release ring to release hose.
Auto Clean interval set to
“OFF”.
Activate cleaner via menu.
Auto Clean interval set to
long interval, time not yet
expired.
Wait for time to expire, or manually
trigger cleaning cycle. Refer to section
21.2.2 Functional Button “ON”,
“Change of cleaning frequency”.
Air flow adjuster too
stringent.
Turn Adjuster on Auto Clean Control
counter-clockwise with a screw driver.
Wiper or Cylinder not
clean.
Air flow adjuster too
wide.
Turn Adjuster on Auto Clean Control
clockwise with a screw driver.
Check wiper and air supply.
Try additional manual cleaning.
Deposition of material
that cannot be removed
mechanically (e.g. Iron
Oxide.)
Try chemical removal of deposit. Apply
Cleaning Kit. Iron Oxide can be
removed with Citric Acid.
Cleaner frequency not
adjusted to situation on
site.
Adjust cleaner frequency.
Wiper seal worn out or
damaged.
Please do not hesitate to contact DECKMA Hamburg for additional support.