Reset (menu Reset) restores factory settings and returns the
Gage to a known condition. The following occurs:
- All batches, stored measurements, batch names and screen
captures are erased.
- Calibration adjustments are returned to factory settings.
- The last stored background measurement (
γ
1
) is erased.
- Menu settings are returned to the following:
1. Power down the Gage and wait 5 seconds.
2. Simultaneously press and hold the and buttons until
the Reset symbol appears.
This returns the Gage to a known, “out-of-the-box” condition.
It
performs the same function as a menu Reset with the addition of:
9
- Date, Time and WiFi settings are not affected by either Reset.
Reset
Setup Menu
Memory = OFF
Bresle Mode = ON
Auto Dim = ON
Display = None
Patch Volume = 3.0 ml
Hi Res = OFF
Bluetooth & Stream = OFF
WiFi & Access Point = OFF
USB Keyboard & Stream = OFF
BLE Keyboard = OFF
Surface Density units = mg/m
2
Temperature units = °C
Units
Perform a more thorough Hard Reset as follows:
Flip Lock = OFF
Backlight = Normal
Auto Sync .net = ON
Sound = Medium
Touch = ON
Language = English
Battery Type = Alkaline
Bluetooth Smart = OFF
USB Drive = ON
- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following status:
NOTE:
Allows the operator to change the displayed units for both
Temperature (°C or °F) and Surface Density (mg/m
2
or µg/cm
2
).