OPERATING INSTRUCTIONS
32 OPERATING INSTRUCTIONS
Leak Test Results (Small Leak):
• If the leak test has failed due to a small leak, a “SMALL
LEAK” notification will be visible
• To initiate a new leak test, touch the RETRY button
• Touch the BACK button to exit the leak test and
return to the Main Screen as necessary
NOTE: This indicates that there is a leak in either the test
cuff, hose, connection points or in the PTS ii internal
pneumatics. Repeat the test with a variety of different
cuffs and hose assemblies. If the unit continues to fail, the
leak is internal, and the PTS ii must be serviced
Leak Test Results (Large Leak):
• If the leak test has failed due to a large leak, a “LARGE
LEAK” notification will be visible
• To initiate a new leak test, touch the RETRY button
• Touch the BACK button to exit the leak test and
return to the Main Screen as necessary
NOTE: This indicates that there is a leak in either the test
cuff, hose, connection points or in the PTS ii internal
pneumatics. Repeat the test with a variety of different
cuffs and hose assemblies. If the unit continues to fail, the
leak is internal, and the PTS ii must be serviced