Problem Resolution
Can’t turn on the handpiece
Please plug-in the power supply and charge the
battery. Then push the ON/OFF button again.
Please contact your qualified technician.
When the battery has malfunctioned and can’t be
charged, the Display will glow and show the sign of
“Er.” Please call your qualified technician to replace
it with a new battery.
This “OH” means the temperature is too high.
Please wait a moment and use it later.
The intensity is too low
If the result of the intensity test in Standard mode is
under 700 mW/cm2, it means the output is too low.
Please contact with your dealer.
6.
Troubleshooting
5.1. During operation, the light should be aimed straightly on the resin to ensure
solidification effectively.
5.2. Never aim the light directly at unprotected soft tissues because it may lead to
injury or irritation. Do not aim the light at eyes. Light reflected from the tooth surface
may also injure eyes. Use the protective shield supplied with the unit or suitable
light filtering safety glasses.
5. Precaution
12