2.7. Maintenance
Prevention of cross infection, cleaning, disinfection and storage.
The use of the sleeve is an additional precautionary measure against
contamination and does not substitute disinfection of the device.
After using, remove the sleeve. Disinfect the light guide and handpiece with
commercial alcohol based surface disinfecting solution. Keep solvents or
ammableliquidsawayfromtheunitcausetheymaydamageitsplastichousing.
Always keep the charger, handpiece and light guide well. The moisture may
cause electrical short-circuit or dangerous malfunction.
1. The curing time interval should exceed 5 seconds
for each cycle.
2. Verify the radiometer sensor which can impact the
accuracy of the measurement. The surface of sensor
area can be wiped with a cotton swab with alcohol.
3. Carefully hold the unit, so the Light Guide
Attachment is flushed with the radiometer sensor
and centered within the white circle provided.
STANDARD
05
Tips:
Optical fiber light guide must be horizontal
alignment in the middle of the sensor area.
Test the Light Guide Attachment with the Radiometer
Verify the LEDEX
TM
WL-090 performance each time before using the radiometer
which is built into the cradle.