EasyManua.ls Logo

DERRICK DE-7200 - VFD Temperature Trend Monitoring; VFD Fault Reset Procedure

Default Icon
167 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATING INSTRUCTIONS
30 Sep 13 4-17
DE-7200VFDCentrifuge
Figure 4-14. Pump VFD Fault Status Screen
VFD TEMPERATURE TRENDS
The historical temperature trends of all three VFDs are shown on the Temperature Trend screen
(Figure 4-15). This screen is accessible from the Bowl VFD screen.
Figure 4-15. VFD Temperature Trend Screen
VFD FAULT RESET
The Fault Reset screen (Figure 4-16) informs the operator of the readiness of each VFDFaulted
or No Faultand permits resetting a faulted VFD. The screen may be selected from either the
Home, Clean Out, or Operation screen. After a bowl or conveyor VFD fault, be sure that the
centrifuge has come to a complete stop before attempting to re-start.

Table of Contents