Important Safeguards
5
WARNING
To reduce the risk of injury:
1. Keep all units away from children. Do not allow unauthorized or
untrained individuals to operate the equipment. Never tamper with or try
to repair the equipment yourself. If you have any questions or suspect
your equipment is not operating properly, contact your oxygen provider.
2.Do not immerse in liquids or subject device to harsh conditions.
3.Do not use in temperatures greater than 104°F (40°C) or below 41°F
(5°C).
4.Do not use with other equipment (i.e. humidifier, nebulizer, etc.) when
in PulseDose delivery mode.
PHYSICIAN’S NOTES
1.Do not use with patients who breathe below 6 Breaths Per Minute
(BPM) or above 40 BPM.
2.Do not use with patients who consistently fail to trigger equipment
(i.e. mouth breathing with closed soft palates).
3.Verify patient is getting adequate Pa02 or Sa02 levels in PulseDose
delivery.
4.Use only standard nasal cannula with PulseDose delivery. Do not use
pediatric (low-flow) nasal cannula or mask with PulseDose delivery.
5. A mask or any nasal cannula can be used with continuous flow delivery.
6. The standard device contains a preset 2 LPM continuous flow backup.
The service manual describes how to change the cannula fitting to
obtain a 3, 4, 5 or 6 LPM continuous flow
backup.
NOTE– This conserving device is designed to fit onto
870-style or “pin-index” cylinders as pictured.
Oxygen
Cylinder
Post Valve
Plastic Tab