Operating Instructions
10
INSERTING A BATTERY INTO THE PULSEDOSE
COMPACT CONSERVING DEVICE
1.Open the battery door.
2.Insert 2 “AA” Alkaline or NiMH batteries.
3.Close the battery door.
NOTE– When changing batteries, first turn the rotary
selector to the “OFF” position.
ATTACHING YOUR PULSEDOSE COMPACT
CONSERVING DEVICE TO THE CYLINDER
1.Loosen the knob.
2.Carefully lower the conserving device over
the post of the cylinder.
CAUTION– Alignment pins can damage sealing surfaces of the post increasing the
chance of leakage.
3.Align the pins in the conserving device to the holes in the cylinder
post as you would a standard regulator.
4.Hand-tighten the knob until the conserving device is secure.
USING YOUR PULSEDOSE COMPACT CONSERVING DEVICE
1.Open the cylinder.
2.Attach the standard nasal cannula to the conserving device and to
your nose and face. Oxygen tubing up to 35 feet long may be used in
PulseDose delivery mode.
3.Turn the rotary selector to the prescribed flow setting.
4.Breathe normally, the conserving device will deliver a bolus of oxygen
at the leading edge of inspiration on every breath up to 40 breaths per
minute.
5.When you are finished using the PulseDose conserving device, turn
your cylinder to the closed position, and rotate the rotary selector to
the “OFF” position.