0044
EN
Drive DeVilbiss® 10-Liter Oxygen
Concentrator Instruction Guide
WARNING–Read instruction guide before operating this equipment.
ASSEMBLED IN USA
NO SMOKING ONLY
ES
Guía de instrucciones del concentrador de
oxígeno de 10-litros de Drive
DeVilbiss®
ADVERTENCIA–Lea la guía de instrucciones antes de poner a funcionar este equipo.
ENSAMBLADO EN EE. UU.
NO FUMAR ÚNICAMENTE
FR
Guide d’instructions du concentrateur
d’oxygène 10-litres Drive
DeVilbiss®
AVERTISSEMENT–Lire le mode d’emploi avant d’utiliser ce dispositif.
ASSEMBLÉ AUX ÉTATS-UNIS
NE PAS FUMER UNIQUEMENT SUR PRESCRIPTION
DE
Drive DeVilbiss® 10-Liter-
Sauerstoffkonzentrator
Bedienungsanleitung
WARNUNG–Vor Inbetriebnahme des Gerätes Bedienungshinweise lesen.
HERGESTELLT IN DEN USA
RAUCHEN VERBOTEN VERSCHREIBUNGSPFLICHTIG
IT
Concentratore di ossigeno da 10-litri Drive
DeVilbiss® Istruzioni per l’uso
AVVERTENZA–Leggere il manuale di istruzioni prima di usarel’apparecchio
ASSEMBLATO NEGLI U.S.A.
NON FUMARE SOLO
NL
Instructiehandleiding Drive DeVilbiss®
10-liter zuurstofconcentrator
WAARSCHUWING–Lees dit instructiehandboekje zorgvuldig door voordat u het
apparaat gaat gebruiken.
GEMONTEERD IN DE VERENIGDE STATEN
VERBODEN TE ROKEN UITSLUITEND OP RECEPT
TR
Drive DeVilbiss® 10-Litre Oksijen
Konsantratörü Kullanım Kılavuzu
UYARI–Cihazı kullanmaya başlamadan önce bu kılavuzu okuyunuz.
ABD’DE MONTE EDILMIŞTIR
SİGARA İÇİLMEZ YALNIZCA REÇETE ILE SATILIR.
PT
Manual de instruções do Concentrador de
oxigênio Drive
DeVilbiss® de 10-litros
ADVERTÊNCIA– Leia o manual de instruções antes de operar este equipamento.
MONTADOS NOS EUA
PROIBIDO FUMAR SOMENTE
PL
Instrukcja obsługi 10-litrowego
koncentratora tlenu Drive
DeVilbiss®
OSTRZEŻENIE – LPrzeczytaj instrukcję obsługi przed rozpoczęciem korzystania z tego
urządzenia.
ZMONTOWANO W STANACH ZJEDNOCZONYCH
NIE PALIĆ WYŁĄCZNIE NA RECEPTĘ

AR
10®DeVilbiss Drive




 
1025 Series
Serie de
1025
Série 1025
1025
Serie
Serie da
1025
1025 Serie
1025 Serisi
Série
1025
Seria 1025
1025


Table of Contents

Questions and Answers

  • B
    brobertsonAug 4, 2025
    What to do if DeVilbiss drive 1025 Oxygen Equipment does not turn on?
    • B
      Benjamin AndersonAug 4, 2025
      If your DeVilbiss Oxygen Equipment isn't turning on (no lights, pulsing alert), first, ensure the power cord is fully plugged into the wall outlet, and on 230V units, check the mains connection at the back. Then, check your home circuit breaker and try a different wall outlet. Finally, press the reset button on the concentrator's circuit breaker, located below the power switch. If issues persist, contact your Drive DeVilbiss provider.
  • E
    Earl WagnerAug 9, 2025
    Why is the red service required light on my DeVilbiss drive 1025 Oxygen Equipment?
    • O
      olsonrandallAug 9, 2025
      If the red service light is on and the alarm is sounding on your DeVilbiss Oxygen Equipment, ensure the flow meter is correctly set. Also, check the air filter and exhaust area for blockages. For the air filter, follow the cleaning instructions. If the above remedies do not work, contact your Drive DeVilbiss provider.
  • I
    ihannaAug 14, 2025
    Why is the yellow low oxygen light on and audible signal sounding on my DeVilbiss drive 1025?
    • E
      Erin LopezAug 14, 2025
      If the yellow low oxygen light is on and the intermittent audible signal is sounding on your DeVilbiss Oxygen Equipment, ensure the flow meter is correctly set to the prescribed number. Also, check the air filter and exhaust area for blockages. If the filter is dirty, follow the cleaning instructions. If the above remedies do not work, contact your Drive DeVilbiss provider.
  • S
    Scott HarrisAug 19, 2025
    What to do if the red service required light is on for DeVilbiss drive 1025 Oxygen Equipment?
    • J
      John KimAug 19, 2025
      If the red service light is illuminated and the audible alert is sounding on your DeVilbiss Oxygen Equipment, check the air filter and wash it if dirty. Also, check the exhaust area to ensure it is not blocked. Inspect the cannula, face mask, and oxygen tubing for blockages or defects, clean or replace if necessary. Detach the humidifier; if proper flow returns, clean or replace it. Finally, ensure the flow meter is set to the prescribed rate. If these steps don't resolve the issue, contact your Drive DeVilbiss provider.
  • A
    Andrea JensenAug 25, 2025
    What to do if DeVilbiss drive 1025 has audible low-frequency vibration?
    • E
      Eileen LutzAug 25, 2025
      If you detect a low-frequency vibration sound from your DeVilbiss Oxygen Equipment, turn the unit off, switch to your reserve oxygen system, and contact your Drive DeVilbiss provider immediately, as this may indicate an electronic assembly malfunction.
  • J
    Jennifer LittleAug 28, 2025
    What to do if other problems occur with DeVilbiss drive 1025 Oxygen Equipment?
    • J
      Jessica Murphy MDAug 29, 2025
      Turn your DeVilbiss Oxygen Equipment “Off.” Switch to your reserve oxygen system and contact your Drive DeVilbiss provider immediately.

Summary

Symbol Definitions

Important Safeguards

Introduction

Intended Use

Defines the primary purpose and application of the oxygen concentrator.

Contraindications

Lists conditions or situations where the device should not be used.

How Your Drive DeVilbiss Oxygen Concentrator Works

Explains the operational mechanism of the oxygen concentrator.

Important Parts of Your Concentrator

Setting Up Your Oxygen Concentrator

Operating Your DeVilbiss Oxygen Concentrator

DeVilbiss OSD Operation

Details the function and indicators of the Oxygen Sensing Device (OSD).

Reserve Oxygen System

Explains the backup oxygen system and its use during power loss or malfunction.

Caring for Your DeVilbiss Oxygen Concentrator

Air Filter

Instructions for cleaning and maintaining the air filter.

Exterior Cabinet

Guidance on cleaning the external surfaces of the concentrator unit.

Troubleshooting

Overview of Alarms

Return and Disposal

Providers Notes - Cleaning and Disinfection When There is a Patient Change

Disinfection

Procedures for disinfecting the concentrator between patient uses.

Specifications

Electromagnetic Compatibility Information

Warranty

Warranty Claims Submissions

Steps to follow for submitting a warranty claim for the equipment.

Ordering and Returning Parts

Information on how to order replacement parts or return components.

DeVilbiss drive 1025 Series Specifications

General IconGeneral
Storage Temperature-4°F to 140°F (-20°C to 60°C)
Operating Humidity15% to 95% non-condensing
Oxygen Concentration87% to 96%
Flow Rate0.5 - 5 LPM
Input Frequency50/60 Hz
Operating Temperature5°C to 40°C