6. Technical specifications
Particle detection range 0.3~10μm
Six-channel particle size 0.30.51.02.55.010um
PM2.5 Range 01000μg/m³
PM2.5 Precision 0~100μg/m³±10μg/
100~1000μg/m³±10%
Concentration resolution g/m³
Number of particles output range ≤ 2000000 Granule/L
Flow 1 liter/min(2.83L=1L*2.83
Cleanliness standard meets the ISO 14644-1 ISO5 level-ISO9 level
Response time 1 second
Boot stabilization time
10 seconds
Temperature measurement range
-1060℃
Temperature accuracy
±2℃
Temperature resolution
0.01℃
Humidity measurement range 0100 RH
Humidity accuracy ±3 RH
Humidity resolution 0.01 RH
Power supply 3.7V Removable 18650 lithium battery
Charging voltage 5V/USB
Length of work 2000mAhCan work continuously≥9 hours
Data record
999 groups
Weight
320g
Size
222*80*46mm
7. Common problem analysis
1. There is no data to show the number and concentration of particulate matter in the air
Analysis 1: The battery power is too low to start the sensor work
2. Doesn't boot
Analysis 1: If the battery is dead or damaged, use the USB port to supply power. If it can be
turned on and there is a change in the power grid in the upper right corner, the battery is dead.
If there is no change, only the OK icon will be displayed.