Test y Ajustes
4
2 CALIBRATION
These functions allow the user to program, adjust and calibrate the scale.
The scale must be turned on for at least two hours before being adjusted. It must also be perfectly
level and the temperature and humidity must be stable. The user must have a calibrated weight relative to
the scale capacity.
The keys have the following functions:
Turn the scale on, and while it is counting down from 9 to 0, press the two keys together, and
; The scale will enter test ready. Next, press the adjustment key found on the CPU. This may be
accessed by removing the sealed screw and inserting a pointed object in the opening.
This should only be done by authorized personnel.
2.1 COUNTRY
Select the country code. The language of message and the rounding will modified according to the country.
To change the country press the key
or
.
Press the key
to program the next field.