17
40.0℃/104.0℉ for at least 30 minutes
before taking a temperature
measurement.
3.If the problem persists after trying
the above method, please to contact
the dealer or manufacturer.
Err
Environment
temperature is not in
10℃ to 40℃ or 50℉
to 104℉ range.
1.Firstly make sure the probe is clean
and no dirty on it.
2.If it`s the enviroment tempertature
out of range, then patient and
thermometer should both stay in a
room between 10.0℃/50.0℉ and
40.0℃/104.0℉ for at least 30 minutes
before taking a temperature
measurement.
3.If the problem persists after trying
the above method, please to contact
the dealer or manufacturer.
9. Specification
Medical infrared Thermometer
Applicable
regulations and laws
ASTM E 1965 / EN12470-5 / GB/T 19146
DC 3.0V (2 pcs of AAA battery)
More than 2000 measurements
Appears on screen when the voltage is below 2.6V±0.1V
℃
℉, c o n v e rtib le
℃/89.6℉ to 42.9℃/109.2℉
℃/0 . 4 ℉
℃/℉
Automatic shutdown
30s±1s
Operation environment
℃
℉
℃
℉
Humidity: 20%-85% RH, non-condensing
Atmospheric pressure:86kPa ~ 106kPa
Storage & shipping
environment
℃
℉
℃
℉
Humidity: 20%-93% RH, non-condensing
Atmospheric pressure:86kPa ~ 106kPa