EasyManuals Logo

Dillon Quantrol AFG Series User Manual

Dillon Quantrol AFG Series
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
Advanced Force Gauge User’s Manual
24
Figure 11
1st Peak Tension and Compression
The display will show % DROP OFF and SET. Press
ENTER to change % DROP OFF to % DROP ON.
Press DOWN to move the arrow cursor to SET % and
press ENTER.
To determine what precisely is considered a break,
you must define the % drop of full-scale value from
the peak load observed prior to the break occurring.
To set the required % drop use UP and DOWN keys
to adjust the value and press ENTER to confirm the
selection.
The % drop value selected also acts as a threshold,
below which the % drop function will not be active.
AFG 100N has % drop of 20 (= 20N). If the peak load
before sample break is 50N, the load must drop to
30N in order for the AFG to detect a 1st peak of 50N.
If load continues to be applied above 50N (e.g. to
75N), the AFG will return 75N as MAX and 50N as 1st
peak.
EXAMPLE
% DROP sub-menu 1
% DROP sub-menu 2
This test would produce
a first peak result.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dillon Quantrol AFG Series and is the answer not in the manual?

Dillon Quantrol AFG Series Specifications

General IconGeneral
BrandDillon
ModelQuantrol AFG Series
CategoryTest Equipment
LanguageEnglish