What to do if there is a major measurement fault with Dräger Aerotest Simultan HP Test Equipment?
- KKaren EdwardsAug 9, 2025
If you're experiencing a major measurement fault with your Dräger Test Equipment, several factors could be the cause. First, verify that the tube is inserted in the correct holder. Additionally, the tube holder might be dirty or loose, so try cleaning or replacing it. Finally, a blocked feed system can cause the wrong volumetric flow. To resolve this, unscrew the tube holder from the measuring unit and use compressed air to clear the feed insert from both sides.