WARNING
Strictly follow the Instructions for Use. User must fully
understand and strictly observe these instructions. The
product must only be used for purposes specified here
.
Dräger Aerotest
HP, HP USA, HP NF, HP NOx, Alpha, Navy, MultiTest med. Int.

Table of Contents

Questions and Answers

  • J
    Jason MeyerAug 8, 2025
    What to do if there is a large measuring error with Dräger Aerotest HP Test Equipment?
    • C
      Chelsey WallaceAug 8, 2025
      If you are getting a large measuring error, there could be a couple of reasons. First, check that the Dräger tube is inserted in the correct Dräger tube holder. Second, the Dräger tube holder might be dirty or loose. Try cleaning the Dräger tube holder, and replace it if necessary.
  • T
    Tracey DillonAug 12, 2025
    Why is the compressed gas connection leaky on my Dräger Aerotest HP Test Equipment?
    • D
      Dr. David BrownAug 12, 2025
      A leaky compressed gas connection is often due to a defective sealing ring. Check the sealing ring and replace it if necessary.
  • J
    Jaclyn OdonnellAug 16, 2025
    What to do if the manual connection cannot be loosened on Dräger Aerotest HP Test Equipment?
    • C
      Chelsey GardnerAug 16, 2025
      If the manual connection is stuck, it may be subject to pressure. Bleed the connection to release the pressure.
  • J
    Jennifer BartonAug 20, 2025
    Why is the safety valve releasing air on my Dräger Test Equipment?
    • M
      Mr. Christopher HooverAug 21, 2025
      If the safety valve is releasing air, it's possible the piston is leaky. Repair the pressure reducer.
  • M
    Mary Sandoval PhDAug 25, 2025
    What to do if the dosing unit is leaky in Dräger Aerotest HP Test Equipment?
    • D
      Derrick HarrisAug 26, 2025
      If the dosing unit is leaky, the seal may be defective. Replace the sealing ring to resolve this issue.
  • H
    hoffmanjulieAug 29, 2025
    What to do if the Dräger tube is not firm in the test holder of my Dräger Aerotest HP Test Equipment?
    • A
      Ashley YoungAug 29, 2025
      If the Dräger tube isn't fitting firmly in the test holder, the Dräger tube container may be worn out. Replace the Dräger tube holder.
  • E
    emmasnowSep 3, 2025
    How to fix a Dräger tube that is not cleanly scored and broken off in Dräger Aerotest HP?
    • N
      Nicholas SimonSep 3, 2025
      If the tip of the Dräger tube isn't cleanly scored and broken off, the break-off unit is likely blunt. Replace the break-off unit.

Dräger Aerotest HP Specifications

General IconGeneral
ApplicationsTesting the purity of compressed air in high-pressure systems
Pressure RangeUp to 300 bar
CalibrationNo calibration required for the device itself; Dräger tubes must be within their expiration date
WeightApprox. 1.5 kg
Detected substancesCO, CO2, Oil mist, Water vapor
Measurement RangeVaries depending on the Dräger test tubes used. Refer to the specific test tube documentation for measurement ranges of individual substances.

Related product manuals