Prepare Dräger tube
Strictly follow the respective Instructions for User for
Dräger tubes.
Break off both tips of the
Dräger tube. Use the Dräger
tube opener for this purpose,
and proceed as follows on
both sides:
5 Insert the Dräger tube as far
as it will go into the hole and
turn it once or twice. The
glass is scored.
6 Tip the Dräger tube to one
side. The tip breaks off and
falls in the break off unit.
5
6
00521765.eps
Use
Explanation using a quadruple measuring device.
The following applies for all Dräger tubes and impactors:
Use Dräger tubes and impactors with the receptacle in the Dräger
tube holder.
Open the valve for the compressed gas supply.
Complete the measurement in accordance with the specifications
in the Instructions for Use for Dräger tubes and impactors.
Close the valve for the compressed gas supply, immediately
remove the Dräger tube from the holder, and use the Instructions
for Use for the Dräger tubes to analyse the results.
First read the Dräger tube "hydro steam 20/a-P".
If the measurement shows that the permitted limits have been
exceeded, it may be necessary to repeat the measurement.
If deviations from the limit values occur again, introduce suitable
measures in the high-pressure system or low-pressure system to
achieve the permitted limit values.
NOTICE
The impactor and the tubes must be used in the tube containers
indicated appropriately, in order for the measurement to be
correct.
When using tubes, ensure the direction of flow is correct.
Only use Dräger tubes, see “Order List” on page 37.
Do not breathe in the gases that occur during the measurement.

Table of Contents