Supplement | Atlan A300, A350 ceiling-mounted versions and wall-mounted version 29
Ceiling-mounted version with service cart
c. If the ceiling supply unit cannot be lowered, use the Up arrow key to raise
the ceiling supply unit sufficiently far that the trunnions on the ceiling supply
unit are pushed into the mount on the device.
d. Press the Down arrow key (12) again.
12.Disconnect Atlan from all devices that are not to be transported.
13.Disconnect the connections between the device and the ceiling supply unit.
a. Disconnect the power supply (13).
b. Disconnect the connection (14) to the central gas supply system.
c. Unplug the system cable (15).
d. Disconnect any other cable connections.
14.Bring the key switch (16) back to its vertical position.