How to clean the Dräger Babyleo TN500
Cleaning in 5 steps
Please complete all cleaning practices according to your hospital‘s guidelines.
This document is for informational purposes only. Refer to the instructions for use
of the device for full disassembly, cleaning and disinfecting instructions.
Prerequisites: There is no patient in the device.
!
1
2
3 4
5
Starting the automatic humidifier cleaning:
• Open the hood.
• On the Main menu touch the
Special procedures
...
• Open the
Humidifier cleaning
tab.
• Touch the
On
button and confirm with
rotary knob.
• Wait until the humidifier cleaning is complete.
This can take up to 60 minutes. Do not
disassemble the device during cleaning.
• Switch off the device and let it cool
before continuing.
Please turn the page for
assembling and dissambling.
Surface disinfection cleaning for:
• Each part removed during disassembly.
• Hood and access panels from inside and outside.
• X-ray flap air ducts (1): clean by pulling a piece
of cloth soaked in disinfectant through each duct.
Machine cleaning and thermal
disinfection for:
• grommets
• fan impeller
Ensure all surfaces are dry before
using for therapy.
Only use disinfectants as listed in the instructions for use.
Make sure no liquid penetrates into the device.
• Display, main column, trolley, handles
and all outer parts.
• Drawer from outside and inside
including insert piece.
!