Instructions for use Babylog 8000 plus SW 5.n 117
Cleaning, disinfection and sterilization
Reprocessing list
Applicable to non-infectious patients.
The reprocessing list contains approximate values
only. The instructions of the hospital’s infection
control officer responsible have priority.
Non-critical medical devices
Semi-critical medical devices
Items which can be repro-
cessed
Recommended repro-
cessing intervals
Manual
Cleaning Disinfection
Ventilator Per patient Yes Yes
Trolley Per patient Yes Yes
Hinged arm Per patient Yes Yes
Flow sensor cable Per patient Yes Yes
Compressed gas hoses Per patient Yes Yes
Items which can
be reprocessed
Recom-
mended repro-
cessing
intervals
Preclean-
ing
Machine
cleaning
and disin-
fection
Manual Steriliza-
tion
Cleaning Disinfec-
tion
Breathing hoses Per
patient/weekly
Yes Yes According to the corre-
sponding instructions
for use
Yes
Y-piece Per
patient/weekly
Yes Yes Possible Possible Yes
Water traps
1)
Water trap con-
tainer
Expiratory valve Per
patient/weekly
2)
Yes Yes Possible Possible Yes
Muffler Per
patient/weekly
Yes Yes Possible Possible Yes
Housing of the
ISO 15 flow sensor
Daily Yes Yes Possible Possible Yes
Y-piece flow sen-
sor
Daily According to the corresponding Instructions for use and section
''Manual reprocessing of the Y-piece'' on page 118