Cleaning, disinfection and sterilization
118 Instructions for use Babylog 8000 plus SW 5.n
Manual reprocessing of the Y-piece
Manual cleaning:
1 Immerse the Y-piece in the solution and agitate
it slightly so that the air can escape.
2 Before the contact time begins and after the
contact time has elapsed:
Insert and remove a pipe cleaner (A) ten
times vertically into each of the two
connection openings of the Y-piece (B) and
then, at an angle, insert and remove the
pipe cleaner ten times into both corners of
the opening for the insert (C).
Fit a syringe (D) containing 20 mL of the
solution to each opening of the Y-piece.
Inject the solution three times.
Perform manual disinfection in the same manner.
Flow sensor insert Daily According to the corresponding Instructions for use
Breathing gas
humidifier
Per
patient/weekly
According to the corresponding instructions for use
Bacterial filter According to the corresponding instructions for use
Medication nebu-
lizer
1)
According to the corresponding instructions for use
Parts for adapting Per
patient/weekly
Yes Yes Possible Possible Yes
1) Keep spring-loaded valves (water trap, pneumatic medication nebulizer) open during reprocessing.
2) Nebulization may lead to increased deposits making it necessary to exchange the parts more often.