EasyManua.ls Logo

Dräger Babylog 8000 plus - Page 118

Dräger Babylog 8000 plus
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cleaning, disinfection and sterilization
118 Instructions for use Babylog 8000 plus SW 5.n
Manual reprocessing of the Y-piece
Manual cleaning:
1 Immerse the Y-piece in the solution and agitate
it slightly so that the air can escape.
2 Before the contact time begins and after the
contact time has elapsed:
Insert and remove a pipe cleaner (A) ten
times vertically into each of the two
connection openings of the Y-piece (B) and
then, at an angle, insert and remove the
pipe cleaner ten times into both corners of
the opening for the insert (C).
Fit a syringe (D) containing 20 mL of the
solution to each opening of the Y-piece.
Inject the solution three times.
Perform manual disinfection in the same manner.
Flow sensor insert Daily According to the corresponding Instructions for use
Breathing gas
humidifier
Per
patient/weekly
According to the corresponding instructions for use
Bacterial filter According to the corresponding instructions for use
Medication nebu-
lizer
1)
According to the corresponding instructions for use
Parts for adapting Per
patient/weekly
Yes Yes Possible Possible Yes
1) Keep spring-loaded valves (water trap, pneumatic medication nebulizer) open during reprocessing.
2) Nebulization may lead to increased deposits making it necessary to exchange the parts more often.
Items which can
be reprocessed
Recom-
mended repro-
cessing
intervals
Preclean-
ing
Machine
cleaning
and disin-
fection
Manual Steriliza-
tion
Cleaning Disinfec-
tion
21887
A
B
C
D

Other manuals for Dräger Babylog 8000 plus

Related product manuals